Yokogawa Insulation Tester MY10-01
Yokogawa Insulation Tester
Yokogawa Portable and Handheld Instrument
MY10 SERIES OF ANALOG TESTERS
Analog models with single rating
MY10-01:125V/20MΩ
MY10-02:250V/50MΩ
MY10-03:500V/100MΩ
MY10-04:500V/1000MΩ
MY10-05:1000V/2000MΩ
AC voltage measurement
Automatic discharge
A wide choice of accessories-Designed for shared use with the MY40
Specifications – Model MY10-01
Rating | 125V/20MΩ |
Effective Measuring Range | 0.01-20MΩ |
Central Scale Value | 0.5MΩ |
AC Voltage Measurement | 0 to 250 V |
Lower Limit | 0.125MΩ |
Rated Current | 1-1.2 mA |
Approx. Overall Scale Length | 107 mm |
Battery Life | Approx. 10 hours when continuously operated on manganese – oxide batteries with the pointer pointing to the central scale value. |
Batteries | Four AA (R6P) batteries |
Approx. Width | 125 mm |
Approx. Height | 103 mm |
Approx. Depth | 53 mm |
Approx. Weight | 500 g |
Compliance | EN61010-1:1993; EN61010-2-31:1995 (Overvoltage Category III, Pollution Degree 2 installations for indoor use) |
Standard Accessories | Batteries (Qty – 4) Earth Probe (98002) (Qty – 1) Line Probe (98001) (Qty – 1) Protection Cover (93013) (Qty – 1) Shoulder Strap (99005) (Qty – 1) User’s Manual (Qty – 1) |
Standard Test Conditions – Model MY10-01
Ambient Temperature/Humidity Ranges | 23 ±5 /45-75% RH |
Position of Use | Horizontal (5° max. of angle of inclination) |
Battery Voltage | Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.) |
Tolerance Mentioned Conditions | Resistance measurement: First effective measuring range = ±5% of reading Second effective measuring range = ±10% of reading Infinite and zero scale values: 0.7% max. of scale length AC voltage: ±10% of maximum scale value No-load voltage: 130% max. of rated voltage Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range Short-circuit current: 12 mA max. |